Table 32. Ventricular tachyarrhythmia therapy parameters (continued)
Parameter
Programmable values
Shipped
Reset
Interval Dec
0; 10 … 40 ms
VF Rx1:
10 ms
Others: —
—
# Sequences
1; 2 … 10
VT Therapies: 3
FVT Therapies: 1
VF Rx1: 1
Others: —
—
On; Off
VF Rx1: On
Others: —
—
Ramp therapy parameters
Initial # Pulses
1; 2 … 8 … 15
—
—
R-S1 Interval=(%RR)
50; 53; 56; 59; 63; 66 … 84; 88;
91 ; 94; 97%
—
—
Interval Dec
0; 10 … 40 ms
—
—
# Sequences
1; 2 … 10
VT Therapies: 3
FVT Therapies: 1
—
—
On; Off
—
—
Ramp+ therapy parameters
Initial # Pulses
1; 2; 3 … 15
—
—
R-S1 Interval=(%RR)
50; 53; 56; 59; 63; 66 … 75 … 84;
88; 91; 94; 97%
—
—
S1S2(Ramp+)=(%RR)
50; 53; 56; 59; 63; 66; 69 … 84; 88;
91; 94; 97%
—
—
S2SN(Ramp+)=(%RR)
50; 53; 56; 59; 63; 66 … 84; 88; 91;
94; 97%
—
—
# Sequences
1; 2 … 10
VT Therapies: 3
FVT Therapies: 1
—
—
On; Off
—
—
Shared Settings…
V-V Minimum ATP Inter-
val
150; 160 … 200 … 400 ms
200 ms
200 ms
V. Amplitude
1; 2 … 6; 8 V
8 V
8 V
V. Pulse Width
0.1; 0.2 … 1.5 ms
1.5 ms
1.5 ms
V. Pace Blanking
150; 160 … 240 … 450 ms
240 ms
240 ms
Active Can/SVC Coil
Can+SVC On ; Can Off; SVC Off
Can+SVC On Can+SVC On
Progressive Episode
Therapies
On; Off
Off
Off
Confi
On ; Off
On
On