B.4 Ventricular tachyarrhythmia therapy parameters
Table 32. Ventricular tachyarrhythmia therapy parameters
Parameter
Programmable values
Shipped
Reset
VF Therapy parameters
VF Therapy Status
On ; Off
On
On
Energy
Rx1–Rx2: 0.4; 0.6 … 1.8; 2; 3 … 16;
18; 20; 22; 24; 25; 26; 28; 30; 32;
35 J
Rx3–Rx6: 10; 11 … 16; 18; 20; 22; 24;
25; 26; 28; 30; 32; 35 J
35 J
35 J
AX>B; B>AX
Rx1–Rx4: B>AX
Rx5–Rx6: AX>B
B>AX
B>AX
ATP…
During Charging ; Before Charging;
Off
During
Charging
Off
Deliver ATP if last 8
R-R >=
200; 210 … 240 … 300 ms
240 ms
240 ms
Therapy Type
Burst ; Ramp; Ramp+
Burst
Burst
ChargeSaver…
On ; Off
On
On
Switch when number
of consecutive ATP
successes equals
1 ; 2; 3; 4; 6; 8; 10
1
4
Smart Mode
On ; Off
On
On
VT/FVT Therapy parameters
VT Therapy Status
On; Off
Off
Off
FVT Therapy Status
On; Off
Off
Off
Therapy Type
CV; Burst; Ramp; Ramp+
Rx1: Burst
Rx2–Rx6: CV
—
—
Energy
0.4; 0.6 … 1.8; 2; 3 … 16; 18; 20; 22;
24; 25; 26; 28; 30; 32; 35 J
VT Rx1–Rx2: 20 J
VT Rx3–Rx6: 35 J
FVT Rx1–Rx6: 35 J
—
—
AX>B; B>AX
Rx1–Rx4: B>AX
Rx5–Rx6: AX>B
—
—
Burst therapy parameters
Initial # Pulses
1; 2 … 8 … 15
VF Rx1: 8
Others: —
—
R-S1 Interval=(%RR)
50; 53; 56; 59; 63; 66 … 84; 88 ; 91;
94; 97%
VF Rx1: 88%
Others: —
—
Medtronic
PROTECTA™ XT DR D314DRG
464
Clinician Manual