5. Observe the Live Rhythm Monitor for the intrinsic rhythm and the match scores for each
compared event. The higher the percentage shown for the match scores, the more
likely the template reflects the patient’s intrinsic morphology.
6. If consistent pacing is still occurring, gradually decrease the Lower Rate to evoke the
intrinsic rhythm.
7. If necessary, you can abort the test by selecting [ABORT]. Pacing settings will return
to the programmed values.
8. After the test completes, you may select [Details] to view details about the stored
template.
10.3.2 Collecting a template
You can use the Wavelet test to manually collect a template if one does not exist, or if the
existing template no longer matches the patient’s intrinsic QRS morphology. Certain factors
such as a change in the patient’s medication or disease progression may have affected the
appropriateness of the currently stored template. See Section 10.3.1 for information about
evaluating the appropriateness of the current template.
Medtronic
PROTECTA™ XT DR D314DRG
Clinician Manual
429