10.2.1 Considerations for measuring pacing thresholds
Selectable and default values – The selectable and default values provided by the Pacing
Threshold Test depend on the programmed values for bradycardia pacing therapy.
Pacing threshold and safety margin – After performing a Pacing Threshold Test, make
sure that the permanently programmed pulse width and amplitude parameters provide an
adequate safety margin above the pacing threshold.
Measuring pacing thresholds in the two-lead system – The device provides
independently selected outputs for Atrial and RV pacing. The Atrial and RV thresholds may
be measured separately and individual safety margins applied to each threshold.
Note: Tachyarrhythmia detection is suspended during a Pacing Threshold Test.
10.2.2 How to measure pacing thresholds
1. Select Tests > Pacing Threshold.
2. Select values for Test Type, Chamber, and Decrement after or accept the values
displayed.
3. Select the starting Test Value for Mode, Lower Rate, AV Delay, Amplitude, and Pulse
Width or accept the values displayed.
4. Press and hold [TEST Press and Hold].
Medtronic
PROTECTA™ XT DR D314DRG
426
Clinician Manual