6.2.2 Information provided by the Quick Look II screen
The Quick Look II screen shows information in 5 sections.
Figure 27. Quick Look II screen
1 Battery information
2 Lead status and trends
3 Pacing and sensing information
4 Arrhythmia episode information
5 Observations
If you select one of the displayed observations and more information about the selected
observation is available, the [>>] button becomes active. You can use the [>>] button to look
at relevant details.
6.2.2.1 Assessing the device and lead status
Battery information – The battery voltage is measured at the start of the session. The
battery voltage is displayed on the Quick Look II screen and is printed in the Initial
Interrogation Report.
Battery voltage is also automatically measured daily at 2:15 AM. If 3 consecutive automatic
daily battery voltage measurements are less than or equal to the Recommended
Replacement Time (RRT) value, the date when the battery reached RRT is displayed along
with the battery voltage on the Quick Look II screen and in the Initial Interrogation Report.
For information on the RRT value, see Section A.2, “Replacement indicators”, page 448.
Medtronic
PROTECTA™ XT DR D314DRG
130
Clinician Manual