AV conduction status. For more information, refer to Section 7.3, “Reducing unnecessary
ventricular pacing with MVP mode”, page 223.
7.15.3 Programming considerations for Mode Switch
MVP modes – Mode Switch is automatically set to On when the pacing mode is set to an
MVP mode (AAIR<=>DDDR or AAI<=>DDD).
Post Mode Switch Overdrive Pacing – You can program Post Mode Switch Overdrive
Pacing (PMOP) to extend pacing in the DDIR mode when the atrial tachyarrhythmia ends.
For more information about PMOP, refer to Section 7.17, “Using atrial intervention pacing
to counteract atrial tachyarrhythmias”, page 277.
7.15.4 Programming Mode Switch
Select Params icon
▷ Mode Switch
7.15.4.1 Programming the Atrial Interval (Rate)
Select Params icon
▷ AT/AF Interval (Rate)
7.15.5 Evaluation of Mode Switch performance
7.15.5.1 EGM strip
Select Data icon
⇒ Clinical Diagnostics
⇒ Arrhythmia Episodes
Select an AT/AF episode from the Arrhythmia Episodes log. Check the A/V bpm column to
evaluate the average atrial and ventricular rates during the episode. Check the EGM column
for an indication that an EGM strip is available for this episode. If EGM is available, select
the EGM option. You can evaluate atrial and ventricular events in the stored EGM strip to
see if the device was operating in a nontracking pacing mode during the episode.