6.8.4.1 Programming data collection preferences
Select Params icon
⇒ Data Collection Setup…
▷ LECG Source
▷ LECG Range
▷ EGM1 Source
▷ EGM1 Range
▷ EGM2 (Wavelet) Source
▷ EGM2 (Wavelet) Range
▷ EGM3 Source
▷ EGM3 Range
▷ Monitored
▷ Pre-arrhythmia EGM
6.9 Viewing episode and therapy counters
The programmer allows you to view stored data about the number of times VT/VF and AT/AF
episodes and therapies have occurred.
The count data for ventricular episodes includes the number of premature ventricular
contractions (PVCs), Ventricular Rate Stabilization (VRS) paces, and monitored and
non-sustained episodes. The count data for ventricular episodes also includes the number
of episodes for which detection and therapy were withheld due to the application of
supraventricular tachycardia (SVT) and ventricular oversensing (VOS) discrimination
features.
The count data for atrial episodes includes the number of monitored, non-sustained, treated,
and pace-terminated episodes.
6.9.1 How to view the counters
Select Data icon
⇒ Clinical Diagnostics
⇒ Counters