Table 56. System test parameters (continued)
Parameter
Selectable values
AV Delay
30; 40 … 350 ms
30; 35 … 60; 70; 75 … 120 bpm
a
The selectable values for this parameter depend on the programmed pacing mode.
b
When performing the test in DDD mode, the Lower Rate must be less than the programmed Upper Tracking
Rate.
c
The selectable values for this parameter depend on the programmed PVAB values.
B.9 EP study parameters
Table 57. T-Shock induction parameters
Parameter
Selectable values
Resume at Deliver
Enabled ; Disabled
Enable
Enabled; Disabled
#S1
2; 3; 4; 5 ; 6; 7; 8
S1S1
300; 310 … 400 … 2000 ms
Delay
20; 30 … 300 … 600 ms
Energy
0.4; 0.6; 0.8; 1.0 … 1.8; 2; 3; 4 … 16; 18; 20; 22; 24;
25; 26; 28; 30; 32; 35 J
Waveform
Monophasic ; Biphasic
AX>B; B>AX
a
If the Active Can/SVC Coil parameter is set to Can Off, the HVA (Can) electrode is not used as part of the high-
voltage delivery pathway. If the Active Can/SVC Coil parameter is set to SVC Off, the HVX (SVC) electrode is
not used as part of the high-voltage delivery pathway.
Table 58. 50 Hz Burst induction parameters
Parameter
Selectable values
Resume at Burst
Enabled ; Disabled
Chamber
RV; Atrium
Amplitude
1; 2; 3; 4 ; 5; 6; 8 V
Pulse Width
0.10; 0.20 … 0.50 … 1.50 ms
VOO Backup (for atrial 50 Hz Burst)
On; Off
Pacing Rate
60; 70 … 120 bpm
V. Amplitude
0.50; 0.75 … 5.00; 5.50; 6.00; 8.00 V
0.10; 0.20 … 1.5 ms
a
The default value for this parameter is set according to the permanently programmed settings for bradycardia
pacing.
b
Crosstalk may occur when atrial pacing amplitude is greater than 6.0 V.