A.5 Magnet application
When a magnet is placed near the device, tachyarrhythmia detection is suspended and no
tachyarrhythmia therapies will be delivered. Alert tones sound if programmed to do so. The
device ignores the magnet in the programmer head when telemetry communication is
established through the programmer head. Before implant and for the first 6 hours after
implant, the device will not sound audible tones when a magnet is placed over the device.
A.6 Stored data and diagnostics
Table 18. Arrhythmia episode data storage
Episode type
Capacity
Treated VT/VF episode log
100 entries
Treated VT/VF episode EGM, markers, and intervals
10 min
Monitored VT episode log
15 entries
Monitored VT episode EGM, markers, and intervals
2.5 min
Non-sustained VT episode log
15 entries
Non-sustained VT episode EGM, markers, and intervals 2 min
High Rate-NS episode log
5 entries
High Rate-NS episode EGM, markers, and intervals
2 min
Treated AT/AF episode log
100 entries
Treated AT/AF episode EGM, markers, and intervals
8.25 min
Monitored AT/AF episode log
50 entries
Monitored AT/AF episode EGM, markers, and intervals
3 min
SVT & V. Oversensing episode log
25 entries
SVT & V. Oversensing episode EGM, markers, and inter-
vals
2.5 min
Rate Drop Response episode log, markers, and intervals 10 entries
Patient activated episode log
50 entries
Flashback memory interval data before each of the fol-
lowing events:
●
Interrogation
●
VT Episode
●
VF Episode
●
AT/AF Episode
2000 events (includes A-A and V-V
events)
Medtronic
PROTECTA™ XT DR D314DRG
Clinician Manual
451