5.8 Replacing a device
Warning: Keep external defibrillation and pacing equipment nearby for immediate use. The
patient does not receive defibrillation or pacing therapy from the device when the lead is
disconnected.
Caution: Disable tachyarrhythmia detection to avoid inappropriate therapy delivery while
explanting the device.
Note: To meet the implant requirements, you may need to reposition or replace the chronic
leads, or add a third high-voltage electrode. For more information, see Section 5.2,
“Selecting and implanting the leads”, page 111.
Note: Any unused leads that remain implanted must be capped with a lead pin cap to avoid
transmitting electrical signals. Contact your Medtronic representative for information about
lead pin caps.
5.8.1 How to explant and replace a device
1. Disable tachyarrhythmia detection to avoid potential inappropriate shocks to the
patient or implanter while explanting the device.
2. Program the device to a mode that is not rate-responsive to avoid potential rate
increases while explanting the device.
3. Dissect the leads and the device free from the surgical pocket. Do not nick or breach
the lead insulation.
4. Use a torque wrench to loosen the setscrews in the connector block.
5. Gently pull the leads out of the connector ports.
6. Evaluate the condition of each lead (see Section 5.3, “Testing the lead system”,
page 113). Replace a lead if the electrical integrity is not acceptable or if the lead
connector pin is pitted or corroded. If you explant the lead, return it to Medtronic for
analysis and disposal.
7. Connect the leads to the replacement device (see Section 5.4, “Connecting the leads
to the device”, page 115).
Note: Lead adaptors may be needed to connect the leads to the replacement device.
Contact a Medtronic representative for information about compatible lead adaptors.
8. Evaluate defibrillation effectiveness using the replacement device (see Section 5.5,
“Performing ventricular defibrillation threshold tests”, page 117).
Medtronic
PROTECTA™ XT DR D314DRG
Clinician Manual
123