Atrial Amplitude and VVI Backup pacing – If the test value for Atrial Amplitude is greater
than 6 V, VVI Backup pacing during a manual atrial ATP therapy may be inhibited by
crosstalk.
Detection suspended during manual therapy – Tachyarrhythmia detection is
automatically suspended when delivering a manual therapy. Detection stays suspended
until you select [RESUME] or the telemetry session between the programmer and device
ends.
Programming head buttons – The Program button on the programming head is disabled
during manual therapies. Use the appropriate on-screen [DELIVER] button to deliver a
manual therapy.
Telemetry – Ensure that a telemetry link is established between the device and the
programmer before you perform a manual therapy. Successful interrogation or
programming confirms proper communication between the device and the programmer.
When using wireless telemetry, verify that at least 3 of the green lights on the wireless
telemetry icon are illuminated. For more information see Section 4.1.
Note: The telemetry link may be lost during the charging period for a high-voltage therapy
due to electrical noise. Telemetry resumes after charging completes.
Temporary parameter values – The manual therapy functions use temporary values that
do not change the programmed parameters of the device. The temporary values take effect
when the manual therapy begins. After the manual therapy, the device reverts to its
programmed parameter values for bradycardia pacing and tachyarrhythmia therapy.
10.8.2 How to deliver a manual therapy
1. Select Tests > EP Study.
2. Select the desired manual therapy from the list of inductions and therapies.
3. If the Select Chamber dialog box appears, select [Atrium] or [RV] as appropriate.
4. Accept the current test values or choose new test values.