8.8 Discriminating RV lead noise from VT/VF
RV lead noise oversensing is caused by lead fracture, breaches in lead insulation, lead
dislodgment, or improper lead connection. If the oversensing is not identified, it may cause
the device to sense non-physiologic noise as rapid ventricular events. If the oversensing is
sustained, such rapid ventricular rates may cause inappropriate delivery of a ventricular
tachyarrhythmia therapy.
8.8.1 System solution: RV Lead Noise Discrimination and RV Lead Noise
Alert
The RV Lead Noise Discrimination feature compares a far-field EGM signal to the near-field
sensing signal to differentiate RV lead noise from VT/VF. If lead noise is identified when
these signals are compared, VT/VF detection and therapy is withheld, and an RV Lead
Noise alert is triggered. This alert issues a tone to warn the patient to contact his or her
clinician, an observation to be viewed by the Medtronic CareLink Model 2090 Programmer,
and a Medtronic CareAlert Notification to warn the clinician that intervention is needed.
8.8.2 Operation of the RV Lead Noise Discrimination feature
When VT or VF is suspected, the device compares the RV sense signal to a far-field EGM
signal, seen on the EGM2 channel. When the RV lead functions properly, these signals
match. If the RV sense signal shows persistent activity in the VT/VF zone that is not shown
on the far-field EGM signal, lead noise is determined, and therapy for VT/VF is withheld.
Note: RV Lead Noise Discrimination is applied only to initial detection.
Figure 175. RV Lead Noise Discrimination
200 ms
T
S
T
S
T
S
F
S
F
S
V
S
F
S
F
S
F
S
F
S
F
S
F
S
F
S
F
S
V
S
V
S
F
S
F
Dl
F
S
F
S
Near-field EGM1 signal
(sensing)
Far-field EGM2 signal
Marker Channel
1 True events are seen simultaneously by the near-field and far-field signals.
2 Noise (shaded area) is indicated by activity on the near-field signal only.
Medtronic
PROTECTA™ XT DR D314DRG
Clinician Manual
345