Sensing – To ensure appropriate delivery of atrial cardioversion therapy, program the
device to prevent sensing of far-field R-waves.
Automatic CV Limits – You can program the device to deliver automatic atrial
cardioversion therapies during selected hours of the day or night. You can also limit the
number of automatic atrial cardioversion therapies that the device can deliver during a single
24-hour cycle.
VF therapy – You must program at least one VF therapy to On before enabling automatic
atrial cardioversion.
9.6.4 Programming atrial cardioversion
Select Params icon
⇒ AT/AF | Therapies…
⇒ AT/AF Rx | Automatic CV…
▷ Automatic CV Status <On>
▷ Energy
▷ Pathway
▷ Minimum R-R Interval
▷ Active Can/SVC Coil
Note: Similar programming steps apply for Automatic CV therapies for Fast AT/AF
episodes.
9.6.5 Evaluation of atrial cardioversion
9.6.5.1 The Quick Look II screen
The Quick Look II screen provides information about AT/AF therapies.
Select Data icon
⇒ Quick Look II
Treated AT/AF episodes – This section includes a count of treated AT/AF episodes. You
can select the Treated [>>] button to view data for the treated episodes.
Quick Look II observations – The Quick Look II observations are based on an analysis
of interrogated data since the last session and programmed parameters. You can select a
specific observation and select the Observations [>>] button to view related information.
Medtronic
PROTECTA™ XT DR D314DRG
412
Clinician Manual