Warning: Make sure that patients understand that they must not carry, store, or leave the
patient magnet positioned over the device. Device operation is temporarily impaired when
the magnet is placed over the device and it must be moved away from the device to restore
normal operation.
●
Instruct patients to contact you immediately if they hear ANY tones from the device.
●
Advise patients of the time of day that you have programmed an alert tone to sound. If
a tone sounds, they should expect it to sound every day at that time until the device is
interrogated.
●
For patients with the CareLink Monitor, advise them that if an alert is not successfully
transmitted to the monitor within 72 hours, a high-urgency tone sounds each day at the
programmed alert time.
●
Make sure patients know that the alert time does not adjust for time zone changes.
●
Advise patients that they may hear a steady test tone or any active alert tones if they
are in the presence of a strong electromagnetic field, such as the field within a store
theft detector. Advise patients that the device operation is temporarily impaired in these
situations, and that they should move away from the source of the interference to restore
normal device operation.
Patients should also understand the purpose of the patient magnet and how and when to
use it. Make sure that they know that current patient alerts sound when the patient magnet
is placed over the device. Demonstrate how to place the patient magnet over the device to
replay the alert tones, and review the patient magnet manual with them. Patients can use a
folded patient magnet manual as a reference card.
6.3.3.5 Demonstrating alert tones
During a patient session, you can demonstrate high-urgency tones, low-urgency tones, and
the test tone from the programmer as follows:
1. Select the Params icon.
2. Select Alert….
3. Select [Demonstrate Tones…].
4. Select [High-Urgency Condition Met], [Low-Urgency Condition Met], or [No Conditions
Met].
Note: Any active alert tones will sound when the patient magnet is placed over the device.
If there are no active alerts, the steady test tone will sound.
Medtronic
PROTECTA™ XT DR D314DRG
Clinician Manual
139