Figure 122. Post Mode Switch Overdrive Pacing (PMOP)
End of AT/AF
episode
Post-Mode Switch
Overdrive Pacing at the
Overdrive Rate
Overdrive Duration
End of mode
switch
A
tri
al
ra
te
min
-1
(bp
m
)
Time (s)
75
90
105
120
135
200
225
0
10
20
40
30
50
Paced beats
Intrinsic beats (atrial tachyarrhythmia)
Post-Mode Switch Overdrive Pacing
Atrial rate
Programmed
pacing rate
AT/AF episode
PMOP is a programmable feature that provides overdrive atrial pacing following the end of
a mode switch. After a mode switch, the device increases the pacing rate beat-by-beat
(decreasing the pacing interval by 70 ms per pulse) until it reaches the programmed
Overdrive Rate. It continues DDIR pacing at the overdrive rate for the duration of the
programmed Overdrive Duration. It then smooths the return to the programmed atrial
tracking mode by gradually slowing the rate (increasing the pacing interval by 70 ms per
pulse) until reaching the programmed pacing rate.
Medtronic
PROTECTA™ XT DR D314DRG
284
Clinician Manual