9.5.2.3 Operation of Burst+ pacing
The programmable parameter Initial #S1 Pulses sets the number of Initial #S1 Pulses in
each Burst+ therapy sequence. A-S1 Interval (%AA), S1-S2 (%AA), and S2-S3 Decrement
are programmable parameters that determine the pacing intervals in a Burst+ sequence.
Each Burst+ sequence consists of the programmed number of Initial #S1 Pulses, followed
by up to 2 additional pulses, if the parameters for these pulses are programmed on. The
pacing intervals for the first Burst+ sequence and additional pulses are determined as
percentages of the atrial tachycardia cycle length. In the first Burst+ sequence, all Initial #S1
Pulses are delivered at the same pacing interval, which is determined by the A-S1 Interval
(%AA) percentage. The first additional pulse is delivered at an interval determined by the
S1-S2 (%AA) percentage. The pacing interval for the subsequent pulse is calculated by
subtracting the S2-S3 Decrement value from the previous interval. This pulse is delivered
only if the S1-S2 (%AA) parameter is programmed on.
If the atrial tachycardia is redetected after an unsuccessful sequence, the device delivers
another Burst+ sequence with shorter pacing intervals. For this sequence, the device
calculates the pacing intervals by subtracting the programmed Interval Decrement value
from each pacing interval in the previous sequence.
VVI ventricular backup pacing is available during Burst+ pacing.
Medtronic
PROTECTA™ XT DR D314DRG
Clinician Manual
399