Notes:
●
When automatic atrial CV therapies become available, they have priority over the
delivery of ATP sequences. After all possible CV therapies have been delivered, the
remaining ATP sequences become available again.
●
If the device detects that the current AT/AF episode has accelerated and become a Fast
AT/AF episode, it skips the remaining sequences of an ATP therapy and starts the next
scheduled therapy for the episode. The device, however, delays therapy for a Fast
AT/AF episode detected after the delivery of an AT/AF pacing therapy. A Fast AT/AF
therapy is delayed for at least 10 min to allow an accelerated rhythm to terminate
spontaneously or revert to the previous AT/AF rhythm.
●
Atrial detection is suspended during the delivery of an atrial ATP therapy sequence.
Figure 204. Overview of atrial ATP therapy delivery
Rx1
Rx2
Sequence 1
Sequence 2
Last sequence
Sequence 1
Sequence 2
Last sequence
Sequence 3
Sequence 3
Rx3
Sequence 1
Sequence 2
Last sequence
Sequence 3
For an overview of atrial ATP sequence delivery, see Figure 205.
9.5.2.1 Atrial ATP therapy scheduling
The device prepares to deliver an atrial ATP therapy if the following conditions are met:
●
An atrial episode is in progress at the time of the scheduled delivery.
●
Atrial ATP therapy sequencing indicates that ATP therapies are enabled for the given
rhythm classification (AT/AF or Fast AT/AF).
●
There is an unused atrial ATP therapy remaining for that classification.
For details about atrial ATP therapy scheduling, refer to Section 9.4, “Scheduling atrial
Medtronic
PROTECTA™ XT DR D314DRG
Clinician Manual
397