![Medtronic PROTECTA XT DR D314DRG Clinician Manual Download Page 235](http://html.mh-extra.com/html/medtronic/protecta-xt-dr-d314drg/protecta-xt-dr-d314drg_clinician-manual_1763148235.webp)
Adjusting the Activity Threshold – For many patients there is no need to reprogram the
Activity Threshold parameter. However, if a patient has minimal rate response during
exercise, the Activity Threshold may need to be programmed to a lower (more sensitive)
setting. The most sensitive setting is “Low”. Conversely, if a patient has an elevated pacing
rate at rest, the Activity Threshold may need to be programmed to a higher (less sensitive)
setting. The least sensitive setting is “High”.
Adjusting Rate Profile Optimization – Before programming other Rate Response
parameters, first verify that the settings for Lower Rate, ADL Rate, and Upper Sensor Rate
are appropriate for the patient.
It may be necessary to reprogram the ADL Response and Exertion Response parameters
if reprogramming the rates does not have the desired effect on Rate Profile Optimization.
By reprogramming the ADL Response and Exertion Response parameters, you can
prescribe a rate profile that matches the patient’s lifestyle or activity levels in each rate range.
Adjust the ADL Response to prescribe how quickly the patient reaches the ADL Rate and
the Exertion Response to prescribe how quickly the patient reaches the Upper Sensor Rate.
In both cases, a lower value decreases the rate responsiveness and a higher value
increases the rate responsiveness.
Note: If increasing the Exertion Response setting does not make Rate Response
aggressive enough, increase the ADL Response setting.
Adjusting the setpoints manually – You can program Rate Profile Optimization to Off
and program the setpoints manually. In this case, the ADL Setpoint and UR Setpoint
determine the pacing rate curve, and rate response calculations continue to operate as
programmed.
7.4.4 Programming Rate Response
Figure 87. Rate Response parameters screens
Medtronic
PROTECTA™ XT DR D314DRG
Clinician Manual