7. If you want to provide VVI Backup pacing during an atrial induction, select values for
VVI Backup.
8. Press and hold [Fixed BURST Press and Hold]. Release the button to end the
induction.
9. If necessary, select [ABORT] to abort a therapy in progress.
10.7.6 Inducing AT or VT with Programmed Electrical Stimulation
You may use Programmed Electrical Stimulation (PES) to induce AT or VT. To induce atrial
or ventricular tachycardias, PES delivers a selectable number of pacing pulses at the S1S1
interval and then delivers up to 3 asynchronous pacing pulses at S1S2, S2S3, and S3S4
intervals. You may specify the chamber, amplitude, pulse width, and pacing intervals for the
induction.
If you perform an atrial PES induction, you may choose to have the device deliver VVI
Backup pacing.
10.7.6.1 Considerations for inducing AT or VT with PES
Warning: Monitor the patient carefully when using an EP study function. Have an external
defibrillator ready for use when inducing any tachyarrhythmia. An induced tachyarrhythmia
may degenerate to ventricular fibrillation.
Atrial Amplitude and VVI Backup pacing – VVI Backup pacing during an atrial PES
induction may be inhibited by crosstalk if the test value for atrial Amplitude is greater than
6 V.
10.7.6.2 How to deliver a PES induction
Figure 224. PES induction screen
Medtronic
PROTECTA™ XT DR D314DRG
Clinician Manual
443