8.8.6.1 Episode Text
Select Data icon
⇒ Clinical Diagnostics
⇒ Arrhythmia Episodes
⇒ Text
For RV lead noise oversensing episodes, the episode text screen offers the following types
of information:
●
an episode summary for V. Oversensing-Noise including:
– Duration
– A/V Max Rate
– V. Median rate
– Activity at onset
●
list of other detection criteria triggered for this episode, and programmed values for all
other detection features
●
description of Wavelet activity if triggered
●
programmed values for VF, FVT, and VT detection and redetection at time of episode
●
status (On/Off) of SVT discrimination features at time of episode
8.8.6.2 Episode EGM
Select Data icon
⇒ Clinical Diagnostics
⇒ Arrhythmia Episodes
⇒ EGM
The Decision Channel annotation “N”, indicating RV Lead Noise Discrimination, is displayed
when the feature withholds ventricular tachyarrhythmia detection. The annotation “NT” is
displayed when RV Lead Noise Discrimination times out.