6.4.5.3 Medtronic CareAlert Events
Select Data Icon
⇒ Alert Events
Check the CareAlert Events list to verify that there is an RV Lead Integrity warning.
Figure 31. CareAlert Event for the RV Lead Integrity warning
6.4.5.4 Sensing Integrity Counter
Select Data Icon
⇒ Device/Lead Diagnostics
⇒ Battery and Lead Measurements
Sensing Integrity Counter – Check the Sensing Integrity Counter section of the Battery
and Lead Measurements screen for evidence of oversensing. For more information, see
Section 6.13, “Viewing detailed device and lead performance data”, page 186.
6.4.5.5 Arrhythmia Episodes (High Rate-NS)
Select Data Icon
⇒ Clinical Diagnostics
⇒ Arrhythmia Episodes
Medtronic
PROTECTA™ XT DR D314DRG
146
Clinician Manual