9.4.5 Evaluation of atrial therapy scheduling
Select Data icon
⇒ Clinical Diagnostics
⇒ Arrhythmia Episodes
⇒ Text
Figure 203. Data – Arrhythmia Episodes screen
For AT/AF episodes, the Arrhythmia Episode text screen lists the following types of
information:
●
an episode summary
●
an event sequence
●
the number of atrial ATP sequences that were delivered in each Reactive ATP region
●
the number of automatic CV shocks that were delivered, if any
●
the number of patient-activated shocks that were delivered, if any
●
the programmed values for AT/AF Detection, Duration to Stop, Reactive ATP,
Automatic CV limits, and the EGM and Sensitivity settings
Medtronic
PROTECTA™ XT DR D314DRG
Clinician Manual