Upon activation, the programmer searches for activated devices within telemetry range. If
a device senses a signal from the programmer, it sends a signal to the programmer and
remains active for 15 min. The programmer then establishes communication with the
device. The Find Patient window is displayed for at least 5 min. If the programmer touch
pen is not used within 5 min, the Find Patient window closes and the Select Model screen
is displayed.
4.1.1.3 How to use the Find Patient window
The programmer lists all patients with wireless-activated implantable devices within
telemetry range in the Patient Name list within the Find Patient window. Patients are listed
in the order in which they were found by the programmer. The list periodically updates to
include patients with newly activated devices and removes patients with devices that are
no longer active, but the order in which the remaining patients appear in the Patient Name
list does not change. The Find Patient window lists patients for whom a session has not
started or has ended.
4.1.1.4 How to verify reliable telemetry between the device and the
programmer
Successful interrogation or programming of the device verifies that reliable telemetry
between the device and the programmer has occurred. When wireless telemetry is first
established during a session, the telemetry status indicator in the upper-left corner of the
programmer task bar changes from the programming head icon to the wireless telemetry
icon shown in Figure 4.
Figure 4. Wireless telemetry icon on the task bar
1 Wireless telemetry icon
The indicator bar on the wireless telemetry icon displays the strength of the wireless
communication signal. Verify that at least 3 of the green lights are illuminated on the indicator
bar to ensure that reliable telemetry has been established between the implanted device
and the programmer.