6.2 Viewing a summary of recently stored data
At the start of a patient session, it is useful to quickly view summary information about device
operation and the patient’s condition over the period since the last follow-up appointment.
This can help you to determine whether you need to look more closely at diagnostic data
or reprogram the device to optimize therapy for the patient.
The Quick Look II screen provides a summary of the most important indicators of the system
operation and patient’s condition. It includes links to more detailed status and diagnostic
information stored in the device. Device and lead status information indicates whether the
system is operating as expected. Information about arrhythmia episodes and therapies
provided gives a picture of the patient’s clinical status since the last follow-up appointment.
System-defined observations alert you to unexpected conditions and suggest how to
optimize the device settings.
Note: The Quick Look II screen shows information collected since the last patient session
and stored in the device memory. Programming changes made during the current session
may also affect the Quick Look II observations.
6.2.1 How to view the Quick Look II screen
The Quick Look II screen is automatically displayed after the patient session is started. You
can also access the Quick Look II screen through the Data icon.
Select Data icon
⇒ Quick Look II
You can update the Quick Look II data during a session by reinterrogating the device.
Medtronic
PROTECTA™ XT DR D314DRG
Clinician Manual
129