AT/AF Therapies Disabled – Atrial therapies can be disabled for the following reasons:
●
A ventricular episode was detected following delivery of an automatic atrial therapy prior
to either redetection of AT/AF or termination of AT/AF. Atrial therapy is disabled if it
appears that an atrial therapy has initiated a ventricular arrhythmia.
●
The Atrial Lead Position Check failed.
●
The device detected an accelerated ventricular rate during ATP therapy.
For more information about disabling atrial therapies, see Section 9.4, “Scheduling atrial
6.14.2 How to respond to the device status indicator warning for electrical
reset
If the programmer reports that an electrical reset occurred and the device is not yet
implanted, do not implant the device. Contact a Medtronic representative. If the device is
implanted, perform the following steps:
1. Remove any sources of electromagnetic interference (EMI).
2. Notify a Medtronic representative.
3. Select [Clear] in the pop-up window to clear the reset indicator and the Medtronic
CareAlert alarm. A confirmation window appears indicating that all previously
interrogated data in the programmer will be cleared.
4. Select [Continue].
5. Interrogate the device.
a. Note the time and date when counter data was last cleared because this indicates
when the electrical reset occurred.
b. Determine, if possible, what the patient was doing at the time and date the electrical
reset occurred.
c. Save your session data. You should give a copy of this saved data file to your
Medtronic representative; it can be helpful in determining the events leading up to
the reset.
6. Verify the programmed device parameters. If a full electrical reset occurred, the primary
reprogrammed values are displayed in the warning message. If a full electrical reset
occurred, reprogram the device parameters.
After this type of reset, the device operates as a simple defibrillator (in VOE-VVI mode)
until it is reprogrammed. For a list of electrical reset parameter settings, see
Appendix B, “Device parameters”, page 459.
Medtronic
PROTECTA™ XT DR D314DRG
196
Clinician Manual