93
Switching On the System
7. Switch on the scan head on the flexible supply unit's main switch board (see
).
8. Switch on the lasers on the main switch board of the flexible supply unit (see
).
The power supplies and fan of the system have now been started.
9. If your system has an infrared laser (MP) or an optical parametric oscillator (OPO),
switch on these components as described in the manufacturer's respective user
manual.
10. If your system has a white light laser, check whether the white light laser's power supply
is switched on: The white light laser's power supply is switched on if the power switch
on the rear side of the white light laser is at "On".
Figure 65:
Power switch on the rear side of the white light laser
11. To switch on the laser in the supply unit, press the key switch on the main switch board
of the flexible supply unit (see
).
WARNING
Risk of permanent eye and skin damage from laser radiation
From this time on, laser radiation may be present in the specimen area of
the laser scanning microscope. Make sure to follow the safety notes for
operation of the system.
12. To switch on the white light laser, activate the detachable key switch on the front side of
the white light laser:
Summary of Contents for TCS SP8 SMD
Page 1: ...10 Living up to Life User Manual Leica TCS SP8 SMD for FCS FLIM and FLCS ...
Page 4: ...4 Copyright ...
Page 14: ...14 Contents ...
Page 18: ...18 Intended Use ...
Page 20: ...20 Liability and Warranty ...
Page 28: ...28 General Safety Notes ...
Page 32: ...32 Additional Notes on Handling the System ...
Page 44: ...44 System Overview and Properties ...
Page 60: ...60 SMD Components Figure 31 DSN 102 Dual SPAD Power Supply ...
Page 80: ...80 Safety Features ...
Page 102: ...102 Switching On the System ...
Page 116: ...116 LAS AF ...
Page 214: ...214 Changing the Specimen ...
Page 216: ...216 Changing the Objective ...
Page 238: ...238 Switching Off the System ...
Page 242: ...242 Repairs and Service Work ...
Page 244: ...244 Maintenance ...
Page 246: ...246 Disassembly and Transport ...
Page 248: ...248 Disposal ...
Page 254: ...254 Contact ...
Page 256: ...256 Recommended Literature ...
Page 266: ...266 Appendix Figure 225 Declaration of conformity ...
Page 268: ...268 Appendix ...
Page 269: ......