193
F(L)CS Data Acquisition
Figure 170:
Settings for using pulsed VIS lasers as excitation source
4. Check whether the correct beam splitter (
MFP
) is selected (
5. You can use the pulsed VIS lasers together with the pulsed UV laser (405 nm). The MFP
setting is not influenced by UV. It is defined by the selection of the pulsed VIS lasers.
18.2.2.4
Using a Pulsed UV Laser
1. Open
the shutter for pulsed lasers.
2. Activate the 405-nm laser by moving the corresponding slider.
3. If no other pulsed laser is used, you can set the multifunction port (
MFP
) to
Substrate
.
Controlling laser intensity of the pulsed laser
For pulsed lasers, the slider does not affect the laser intensity. Laser
intensity needs to be controlled at the desired laser.
Combination pulsed and continuous wave VIS excitation
Do not use any pulsed VIS excitation together with continuous wave VIS
excitation for continuous wave laser lines close to the wavelengths of
the pulsed laser (such as the combination 470 nm 488 nm
continuous wave), as no optimum MFP setting exists for such a
combination.
Summary of Contents for TCS SP8 SMD
Page 1: ...10 Living up to Life User Manual Leica TCS SP8 SMD for FCS FLIM and FLCS ...
Page 4: ...4 Copyright ...
Page 14: ...14 Contents ...
Page 18: ...18 Intended Use ...
Page 20: ...20 Liability and Warranty ...
Page 28: ...28 General Safety Notes ...
Page 32: ...32 Additional Notes on Handling the System ...
Page 44: ...44 System Overview and Properties ...
Page 60: ...60 SMD Components Figure 31 DSN 102 Dual SPAD Power Supply ...
Page 80: ...80 Safety Features ...
Page 102: ...102 Switching On the System ...
Page 116: ...116 LAS AF ...
Page 214: ...214 Changing the Specimen ...
Page 216: ...216 Changing the Objective ...
Page 238: ...238 Switching Off the System ...
Page 242: ...242 Repairs and Service Work ...
Page 244: ...244 Maintenance ...
Page 246: ...246 Disassembly and Transport ...
Page 248: ...248 Disposal ...
Page 254: ...254 Contact ...
Page 256: ...256 Recommended Literature ...
Page 266: ...266 Appendix Figure 225 Declaration of conformity ...
Page 268: ...268 Appendix ...
Page 269: ......