Summarized manual for FCS or other point measurements
7. The settings in the
Measurement Preview
window in SymPhoTime can be preset by
selecting
Show Measurement Preview...
when no measurement is running. Defaults can
be saved using the
Set Defaults
button.
8. The number of calculated points per trace can be adjusted with
NSub
. The channel
assignment is already preset, as shown in
. You should not change this
assignment.
Figure 182:
Online monitoring of the FCS trace and photon count rate in the SymPhoTime preview window.
Left: FCS traces after 60 seconds of acquisition, cross and autocorrelation. G(0) near one
corresponds to an average number of one dye molecule in the confocal detection volume.
Right: Display of the count rate over time. The fluorescence bursts of single molecules are
clearly visible.
19.2
Starting Point/ FCS Data Acquisition
1. Start the FCS Wizard in LAS AF.
2. Scan an image according to the
Setup Imaging
3. Optimize the FCS settings in
step
Setup FCS
and
).
4. Go to the
Measurements
step (
).
5. Select the positions where you want to carry out a point measurement.
'Fifo overrun' error message
At very high count rates, depending on the computer power and number
of calculated points per FCS trace, a
Fifo overrun
error can occur. In this
case, use low Nsub values or switch off the FCS online correlation by
unchecking the
FCS
check box under
Previews
.
Summary of Contents for TCS SP8 SMD
Page 1: ...10 Living up to Life User Manual Leica TCS SP8 SMD for FCS FLIM and FLCS ...
Page 4: ...4 Copyright ...
Page 14: ...14 Contents ...
Page 18: ...18 Intended Use ...
Page 20: ...20 Liability and Warranty ...
Page 28: ...28 General Safety Notes ...
Page 32: ...32 Additional Notes on Handling the System ...
Page 44: ...44 System Overview and Properties ...
Page 60: ...60 SMD Components Figure 31 DSN 102 Dual SPAD Power Supply ...
Page 80: ...80 Safety Features ...
Page 102: ...102 Switching On the System ...
Page 116: ...116 LAS AF ...
Page 214: ...214 Changing the Specimen ...
Page 216: ...216 Changing the Objective ...
Page 238: ...238 Switching Off the System ...
Page 242: ...242 Repairs and Service Work ...
Page 244: ...244 Maintenance ...
Page 246: ...246 Disassembly and Transport ...
Page 248: ...248 Disposal ...
Page 254: ...254 Contact ...
Page 256: ...256 Recommended Literature ...
Page 266: ...266 Appendix Figure 225 Declaration of conformity ...
Page 268: ...268 Appendix ...
Page 269: ......