46
Ambient Conditions
8.3
Room Dimensions
The room should be as large as possible so that multiple operators can find room around the
system.The free space to the right, left and rear of the system should be 60 cm. In front of
the system there must be 110 cm of free space so that the service technicians can access
the system as necessary.
WARNING
Fire or overheating of the system possible
Electrical peripheral devices must be placed at least 10 cm (4") away
from the wall and from flammable substances to prevent overheating or
fire of the system.
Also make sure that doors and passages are at least 100 cm wide in order to be able to
safely transport the system.
The following minimum size should be planned for the space:
Room dimensions (length x depth)
TCS SP8 SMD with upright microscope,
compact and flexible supply unit
possible
470 x 290 cm (15'5" x 9'6")
TCS SP8 SMD with inverted
microscope, compact and flexible
supply unit possible
484 x 290 cm (15'11" x 9'6")
TCS SP8 SMD with white light laser,
upright microscope and PicoQuant
laser
520 x 290 cm (17'1" x 9'6")
TCS SP8 SMD with white light laser,
upright microscope without PicoQuant
laser
470 x 290 cm (15'5" x 9'6")
TCS SP8 SMD with white light laser,
inverted microscope and PicoQuant
laser
534 x 290 cm (17'6" x 9'6")
TCS SP8 SMD with white light laser,
inverted microscope without PicoQuant
laser
484 x 290 cm (15'11" x 9'6")
TCS SP8 SMD with MP configuration
610 x 290 cm (20' x 9'6")
TCS SP8 SMD with white light laser and
MP configuration, without PicoQuant
laser
610 x 290 cm (20' x 9'6")
TCS SP8 SMD with white light laser, MP
configuration and PicoQuant laser
660 x 290 cm (21'8" x 9'6")
Summary of Contents for TCS SP8 SMD
Page 1: ...10 Living up to Life User Manual Leica TCS SP8 SMD for FCS FLIM and FLCS ...
Page 4: ...4 Copyright ...
Page 14: ...14 Contents ...
Page 18: ...18 Intended Use ...
Page 20: ...20 Liability and Warranty ...
Page 28: ...28 General Safety Notes ...
Page 32: ...32 Additional Notes on Handling the System ...
Page 44: ...44 System Overview and Properties ...
Page 60: ...60 SMD Components Figure 31 DSN 102 Dual SPAD Power Supply ...
Page 80: ...80 Safety Features ...
Page 102: ...102 Switching On the System ...
Page 116: ...116 LAS AF ...
Page 214: ...214 Changing the Specimen ...
Page 216: ...216 Changing the Objective ...
Page 238: ...238 Switching Off the System ...
Page 242: ...242 Repairs and Service Work ...
Page 244: ...244 Maintenance ...
Page 246: ...246 Disassembly and Transport ...
Page 248: ...248 Disposal ...
Page 254: ...254 Contact ...
Page 256: ...256 Recommended Literature ...
Page 266: ...266 Appendix Figure 225 Declaration of conformity ...
Page 268: ...268 Appendix ...
Page 269: ......