149
FLIM Data Acquisition
5. If you want to view an online lifetime histogram, close the fast FLIM image in
SymPhoTime. Click the symbol
Show Oscilloscope...
and select the
TCSPC histogram
tab.
6. To cancel specimen illumination, select the
Stop FLIM Test
button in the LAS AF
software. This will also stop the online FLIM display in SymPhoTime.
The instrument parameters set for the FLIM measurements (laser lines and associated
intensity, pinhole, scan mode, image size etc.) are kept as separate FLIM settings that are
independent from the image acquisition setting defined in the
Setup Imaging
step. The FLIM
settings defined in the
Setup FLIM
step are automatically transferred to the
Measurements
step.
16.2.9
Count Rate Monitor
The count rate of each channel is displayed in the
Setup
tab. For FLIM, the count rate should
be between 10,000 and 1,000,000 cps (counts per second).
Figure 135:
Count rate monitor in the FLIM Wizard
In addition, the count rate monitor shows the number of photons in the brightest pixel of the
image. This value allows you to evaluate the image quality. The higher the number of
fluorescence photons (counts) per pixel, the more accurately the lifetime can be
determined. This value is updated regularly whenever a new scanning operation starts in
the
Setup FLIM
step of the FLIM Wizard.
16.2.10
Loading and Saving FLIM-specific Instrument Parameter Settings
The instrument parameter setting (IPS) for FLIM can be saved and loaded in the same way
as IPSs for normal intensity imaging.
Save the FLIM settings in the
Setup FLIM
step and settings for intensity image acquisition in
the
Setup Imaging
step.
Summary of Contents for TCS SP8 SMD
Page 1: ...10 Living up to Life User Manual Leica TCS SP8 SMD for FCS FLIM and FLCS ...
Page 4: ...4 Copyright ...
Page 14: ...14 Contents ...
Page 18: ...18 Intended Use ...
Page 20: ...20 Liability and Warranty ...
Page 28: ...28 General Safety Notes ...
Page 32: ...32 Additional Notes on Handling the System ...
Page 44: ...44 System Overview and Properties ...
Page 60: ...60 SMD Components Figure 31 DSN 102 Dual SPAD Power Supply ...
Page 80: ...80 Safety Features ...
Page 102: ...102 Switching On the System ...
Page 116: ...116 LAS AF ...
Page 214: ...214 Changing the Specimen ...
Page 216: ...216 Changing the Objective ...
Page 238: ...238 Switching Off the System ...
Page 242: ...242 Repairs and Service Work ...
Page 244: ...244 Maintenance ...
Page 246: ...246 Disassembly and Transport ...
Page 248: ...248 Disposal ...
Page 254: ...254 Contact ...
Page 256: ...256 Recommended Literature ...
Page 266: ...266 Appendix Figure 225 Declaration of conformity ...
Page 268: ...268 Appendix ...
Page 269: ......