Summarized manual for FCS or other point measurements
6. Select the measurement times.
7. Start the FCS measurement in LAS AF by selecting the image acquisition button
Run FCS
under the step
Measurements
.
Normally, the SymPhoTime software will start the data acquisition automatically. If you
start the FCS data acquisition by selecting
Acquire Point Measurements
in the
SymPhoTime software, the measurement is not synchronized with LAS AF. The file
name, comments and instrument parameter settings are not transferred from LAS AF to
SymPhoTime.
8. The Measurement Preview dialog opens in SymPhoTime on the
FCS
tab (
left
).
The FCS curve is integrated over time.
Figure 183:
Working area of the annotation file
9. At the end of the measurement, the result can be viewed in the annotation file which can
be opened from the workspace (
). Open the
Correlation
tab as shown in
. The curves can be analyzed by selecting the
FIT
button that corresponds to
the cross-correlation or autocorrelation curve.
10. Starting from the acquired raw data file (in this case,
FCS 01.pt3
), the FCS curve can be
recalculated with enhanced functions, such as Time Gated FCS or Fluorescence Lifetime
Correlation Spectroscopy (FLCS) using the
FCS trace
button. The prerequisite for both of
these modes is that the FCS measurement was carried out using pulsed lasers for
excitation.
Summary of Contents for TCS SP8 SMD
Page 1: ...10 Living up to Life User Manual Leica TCS SP8 SMD for FCS FLIM and FLCS ...
Page 4: ...4 Copyright ...
Page 14: ...14 Contents ...
Page 18: ...18 Intended Use ...
Page 20: ...20 Liability and Warranty ...
Page 28: ...28 General Safety Notes ...
Page 32: ...32 Additional Notes on Handling the System ...
Page 44: ...44 System Overview and Properties ...
Page 60: ...60 SMD Components Figure 31 DSN 102 Dual SPAD Power Supply ...
Page 80: ...80 Safety Features ...
Page 102: ...102 Switching On the System ...
Page 116: ...116 LAS AF ...
Page 214: ...214 Changing the Specimen ...
Page 216: ...216 Changing the Objective ...
Page 238: ...238 Switching Off the System ...
Page 242: ...242 Repairs and Service Work ...
Page 244: ...244 Maintenance ...
Page 246: ...246 Disassembly and Transport ...
Page 248: ...248 Disposal ...
Page 254: ...254 Contact ...
Page 256: ...256 Recommended Literature ...
Page 266: ...266 Appendix Figure 225 Declaration of conformity ...
Page 268: ...268 Appendix ...
Page 269: ......