Summarized manual for FLIM experiments
17
Summarized Manual for FLIM Experiments
17.1
Prerequisites
The following description assumes that the following prerequisites are met:
• SymPhoTime has been started and a workspace has been loaded.
• LAS AF was started.
• All components (lasers, detectors, software) of the PicoQuant SMD upgrade are
switched on. Detailed information on starting the components can be found in
.
• The right objective is chosen with high numerical aperture and - if present - the
correction ring is adjusted to the specimen used (
).
• Your specimen is already placed onto the microscope and brought into focus.
• The selected excitation lasers are switched on and activated (
).
• The correct emission filter(s) was placed in front of the detectors. For external FLIM
detectors, the correct SMD filter cube has to be set in the X1 port adapter (
);
for internal SP FLIM detection, the barrier filter in the fluorifier disc is used
(
). With HyD RLD, the correct barrier filter and a suitable filter cube must
be used in the detector unit.
• For beam splitter systems: The correct beam splitter for your wavelength has been
selected in LAS AF. If pulsed diode lasers are used, the beam splitter is to be in the
Substrate
position.
• With AOBS systems: The acousto-optical beam splitter is in fluorescence mode.
17.2
Selecting Position for the FLIM Measurement
1. Start the Wizard in LAS AF by selecting the
FLIM
operating mode.
2. Scan an image (for this, use the
Setup Imaging
step in the FLIM Wizard). To be able to
get an initial overview, you can use continuous wave lasers and internal detectors.
3. Select the correct sample region at which the FLIM measurement is to be carried out.
17.3
Changing from Continuous to Pulsed Excitation
1. Make sure that the PicoQuant laser driver is set to external trigger.
2. In the FLIM Wizard in LAS AF, go to the
Setup FLIM
step.
3. Select a suitable laser wavelength for pulsed excitation (
).
4. Check whether the selectable beam splitter for
MFP
is in the correct position
5. Select the laser pulse frequency (repetition rate) in LAS AF (
). In general,
40 MHz is the best frequency to start with. Lower the frequency to 20 MHz only if you
determine that your lifetime decay curve does not fit in the TCSPC histogram. This
Summary of Contents for TCS SP8 SMD
Page 1: ...10 Living up to Life User Manual Leica TCS SP8 SMD for FCS FLIM and FLCS ...
Page 4: ...4 Copyright ...
Page 14: ...14 Contents ...
Page 18: ...18 Intended Use ...
Page 20: ...20 Liability and Warranty ...
Page 28: ...28 General Safety Notes ...
Page 32: ...32 Additional Notes on Handling the System ...
Page 44: ...44 System Overview and Properties ...
Page 60: ...60 SMD Components Figure 31 DSN 102 Dual SPAD Power Supply ...
Page 80: ...80 Safety Features ...
Page 102: ...102 Switching On the System ...
Page 116: ...116 LAS AF ...
Page 214: ...214 Changing the Specimen ...
Page 216: ...216 Changing the Objective ...
Page 238: ...238 Switching Off the System ...
Page 242: ...242 Repairs and Service Work ...
Page 244: ...244 Maintenance ...
Page 246: ...246 Disassembly and Transport ...
Page 248: ...248 Disposal ...
Page 254: ...254 Contact ...
Page 256: ...256 Recommended Literature ...
Page 266: ...266 Appendix Figure 225 Declaration of conformity ...
Page 268: ...268 Appendix ...
Page 269: ......