252
Troubleshooting
31.4
How to Handle PQ Error Codes in LAS AF
Error messages generated in SymPhoTime are also displayed in LAS AF. The following table
provides a list of possible messages and recommended user actions.
Table 23: SymPhoTime error codes displayed in LAS AF and recommended measures
PQ error code displayed in LAS AF
Recommended user action
PQ_ERRCODE_NO_ERROR
Continue
PQ_ERRCODE_MEASUREMENT_READ
Y
Retry
PQ_ERRCODE_USER_BREAK
Measurement was stopped by user
-> Retry
PQ_ERRCODE_MESSAGE_CORRUPTED
=-1
Network communication problem
-> Retry, report to Leica
PQ_ERRCODE_SERVER_BUSY = -2
Retry or restart system
PQ_ERRCODE_MESSAGE_TIMEOUT = -
3
Retry
PQ_ERRCODE_INVALID_REC_VERSION
=-10
SymPhoTime and LAS AF version incompatible ->
Latest software versions installed?
PQ_ERRCODE_MEASUREMENT_TIME
OUT =-100
Retry
PQ_ERRCODE_FIFO_OVERRUN = -101
Count rates are too high -> Lower laser intensity
PQ_ERRCODE_DMA_ERROR = -102
Retry
PQ_ERRCODE_OSCILLOSCOPE_RUNNI
NG = -103
Stop oscilloscope measurement in SymPhoTime
PQ_ERRCODE_HARDWARE_INIT = -104
Restart PicoHarp and SymPhoTime computer
PQ_ERRCODE_TTTR_INIT = -105
Restart PicoHarp and SymPhoTime computer
PQ_ERRCODE_TTTR_RUNNING = -106
Either another measurement is active -> Stop
measurement in SymPhoTime
or
SymPhoTime is busy with saving data -> In an FCS or
FLIM time series, increase the time between
individual measurements
PQ_ERRCODE_NO_WORKSPACE = -107
Open workspace in SymPhoTime
PQ_ERRCODE_FILE_EXISTS = -108
Choose new file name in LAS AF that does not appear
in the SymPhoTime workspace
PQ_ERRCODE_FILE_CREATE = -109
Folder is write-protected, choose other folder for
workspace in SymPhoTime
PQ_ERRCODE_GROUPNAME_TOO_LO
NG =-110
Group name must be not longer than 63 characters ->
Choose shorter base name in LAS AF
PQ_ERRCODE_FILENAME_TOO_LONG
=-111
File name must be no longer than 255 characters ->
Choose shorter base name in LAS AF
PQ_ERRCODE_TIMESTAMP_ARRAY_T
OO_LONG =-112
Reduce frame size of FLIM image in x. Contact Leica
PQ_ERRCODE_INVALID_LICENSE =-
999
The software license is not valid for this kind of
measurement
PQ_ERRCODE_UNKNOWN_ERROR = -
9999
Retry
Summary of Contents for TCS SP8 SMD
Page 1: ...10 Living up to Life User Manual Leica TCS SP8 SMD for FCS FLIM and FLCS ...
Page 4: ...4 Copyright ...
Page 14: ...14 Contents ...
Page 18: ...18 Intended Use ...
Page 20: ...20 Liability and Warranty ...
Page 28: ...28 General Safety Notes ...
Page 32: ...32 Additional Notes on Handling the System ...
Page 44: ...44 System Overview and Properties ...
Page 60: ...60 SMD Components Figure 31 DSN 102 Dual SPAD Power Supply ...
Page 80: ...80 Safety Features ...
Page 102: ...102 Switching On the System ...
Page 116: ...116 LAS AF ...
Page 214: ...214 Changing the Specimen ...
Page 216: ...216 Changing the Objective ...
Page 238: ...238 Switching Off the System ...
Page 242: ...242 Repairs and Service Work ...
Page 244: ...244 Maintenance ...
Page 246: ...246 Disassembly and Transport ...
Page 248: ...248 Disposal ...
Page 254: ...254 Contact ...
Page 256: ...256 Recommended Literature ...
Page 266: ...266 Appendix Figure 225 Declaration of conformity ...
Page 268: ...268 Appendix ...
Page 269: ......