212
Summarized manual for FCS or other point measurements
Figure 184:
Display of the FCS curves in the "Correlation" tab of the annotation file
Using detectors with high quantum efficiency
FCS measurements can only be carried out with detectors with high
quantum efficiency, such as APDs. Using cross-correlation between two
detectors,
detector after pulsing
effects can be suppressed all the way
to the correlation curve. With pulsed excitation, alternatively, FLCS can
be used to
calculate detector after-pulses for the individual detection
channels from the raw data.
A stable count rate for FCS analysis is necessary
For a reliable FCS analysis, the count rate has to be stable. Signal
attenuation due to photobleaching may lead to an increasing FCS
correlation amplitude at long lag times.
Summary of Contents for TCS SP8 SMD
Page 1: ...10 Living up to Life User Manual Leica TCS SP8 SMD for FCS FLIM and FLCS ...
Page 4: ...4 Copyright ...
Page 14: ...14 Contents ...
Page 18: ...18 Intended Use ...
Page 20: ...20 Liability and Warranty ...
Page 28: ...28 General Safety Notes ...
Page 32: ...32 Additional Notes on Handling the System ...
Page 44: ...44 System Overview and Properties ...
Page 60: ...60 SMD Components Figure 31 DSN 102 Dual SPAD Power Supply ...
Page 80: ...80 Safety Features ...
Page 102: ...102 Switching On the System ...
Page 116: ...116 LAS AF ...
Page 214: ...214 Changing the Specimen ...
Page 216: ...216 Changing the Objective ...
Page 238: ...238 Switching Off the System ...
Page 242: ...242 Repairs and Service Work ...
Page 244: ...244 Maintenance ...
Page 246: ...246 Disassembly and Transport ...
Page 248: ...248 Disposal ...
Page 254: ...254 Contact ...
Page 256: ...256 Recommended Literature ...
Page 266: ...266 Appendix Figure 225 Declaration of conformity ...
Page 268: ...268 Appendix ...
Page 269: ......