![Leica TCS SP8 SMD User Manual Download Page 198](http://html1.mh-extra.com/html/leica/tcs-sp8-smd/tcs-sp8-smd_user-manual_1873063198.webp)
198
F(L)CS Data Acquisition
18.3.5
Optimizing FCS Settings
1. First, use the image display to find a measuring point in the image. Move the crosshair in
the image display to the desired position (
Figure 175:
Selecting the measuring points
2. Press the
Run FCS Test
button (
) on the
Setup
tab to start the
illumination. You can view and optimize the fluorescence intensity in the count rate
monitor. Laser line and intensity can be modified during the test measurement until you
attain the count rate or counts per molecule desired.
3. The scanner cannot be moved out of its defined position while
Run FCS Test
is still active.
To move the scanner to another position, you first have to stop the FCS measurement by
clicking the
Stop FCS Test
button.
4. Then, move the crosshair to the desired position and restart the measurement by clicking
the
Run FCS Test
button.
When you start the test measurement, an online correlation curve is displayed in
SymPhoTime.
A scanning operation is required before an FCS measurement
After starting LAS AF, at least one scanning operation has to be carried
out (using
Live
or
Capture
) before an FCS measurement can be started.
This is also necessary when working in solutions only. Otherwise, no FCS
measurement is started.
Summary of Contents for TCS SP8 SMD
Page 1: ...10 Living up to Life User Manual Leica TCS SP8 SMD for FCS FLIM and FLCS ...
Page 4: ...4 Copyright ...
Page 14: ...14 Contents ...
Page 18: ...18 Intended Use ...
Page 20: ...20 Liability and Warranty ...
Page 28: ...28 General Safety Notes ...
Page 32: ...32 Additional Notes on Handling the System ...
Page 44: ...44 System Overview and Properties ...
Page 60: ...60 SMD Components Figure 31 DSN 102 Dual SPAD Power Supply ...
Page 80: ...80 Safety Features ...
Page 102: ...102 Switching On the System ...
Page 116: ...116 LAS AF ...
Page 214: ...214 Changing the Specimen ...
Page 216: ...216 Changing the Objective ...
Page 238: ...238 Switching Off the System ...
Page 242: ...242 Repairs and Service Work ...
Page 244: ...244 Maintenance ...
Page 246: ...246 Disassembly and Transport ...
Page 248: ...248 Disposal ...
Page 254: ...254 Contact ...
Page 256: ...256 Recommended Literature ...
Page 266: ...266 Appendix Figure 225 Declaration of conformity ...
Page 268: ...268 Appendix ...
Page 269: ......