Summarized manual for FLIM experiments
17.10
Starting FLIM Data Acquisition
1. Go to the
Measurements
step in the FLIM Wizard in LAS AF.
2. In the
Setup
tab, type in a file name that will be used in SPT (
).
3. Specify the measurement time of the FLIM image (
).
4. If you wish, define a volume or lambda stack or a time series (
).
5. Start the scan procedure in LAS AF by selecting the
Run FLIM
image acquisition button.
Normally, the SymPhoTime software will start the data acquisition automatically.
However, for a single FLIM measurement, you can also start the FLIM data acquisition
manually by selecting the
Acquire LSM Measurements
button.
During the SP8 data acquisition, the preview window is displayed in the SymPhoTime
software (
), in which you can see the FLIM preview image. When scanning the
specimen several times, all frames are accumulated over time. This display allows you to
evaluate the quality of the FLIM data acquired. In doing so, note that the FLIM preview image
is accumulated for display only. Based on the stored raw data, separate individual images
can still be analyzed.
6. The SP8 scanning operation stops either after having scanned the chosen number of
frames or can be stopped manually by selecting
Stop FLIM
to interrupt the data
acquisition.
The recorded FLIM images will be stored in the SymPhoTime software and are accessible
for further data analysis from the workspace.
17.11
Resulting Raw Data File and Documentation
The resulting raw data file and documentation is shown in the SymPhoTime software.
The measurement result is an FLIM image with a *.PT3 raw data file, which contains the
greatest available data volume for each photon detected (temporal and spatial information
related to the detection channel).
Each raw data file is saved with an annotation file. It is characterized in the working area
overview by a symbol in the shape of a red dot (
). The raw data file contains all
measurement parameters transferred to SymPhoTime software and contains a preview
image (bitmap) which provides a first impression of the image scanned.
Prevent light incidence during MP FLIM measurement
During an MP FLIM measurement, the room light should be switched off
and the shutter of the fluorescence lamp should be closed. Furthermore,
the iris diaphragm on the condenser should be completely closed.
Summary of Contents for TCS SP8 SMD
Page 1: ...10 Living up to Life User Manual Leica TCS SP8 SMD for FCS FLIM and FLCS ...
Page 4: ...4 Copyright ...
Page 14: ...14 Contents ...
Page 18: ...18 Intended Use ...
Page 20: ...20 Liability and Warranty ...
Page 28: ...28 General Safety Notes ...
Page 32: ...32 Additional Notes on Handling the System ...
Page 44: ...44 System Overview and Properties ...
Page 60: ...60 SMD Components Figure 31 DSN 102 Dual SPAD Power Supply ...
Page 80: ...80 Safety Features ...
Page 102: ...102 Switching On the System ...
Page 116: ...116 LAS AF ...
Page 214: ...214 Changing the Specimen ...
Page 216: ...216 Changing the Objective ...
Page 238: ...238 Switching Off the System ...
Page 242: ...242 Repairs and Service Work ...
Page 244: ...244 Maintenance ...
Page 246: ...246 Disassembly and Transport ...
Page 248: ...248 Disposal ...
Page 254: ...254 Contact ...
Page 256: ...256 Recommended Literature ...
Page 266: ...266 Appendix Figure 225 Declaration of conformity ...
Page 268: ...268 Appendix ...
Page 269: ......