101
Switching On the System
power switch is in the "ON" position.
3. Start the laser driver and switch the laser on. If you are using the PDL 828 Sepia II laser
driver, set the key switch to the "ON" position. Prior to data acquisition, the laser head
should be operated for a few minutes to stabilize. If you use a PDL 828 Sepia II, start the
specific Sepia software on the SMD workstation and set the laser settings there (refer
to corresponding manuals).
4. Start the SymPhoTime software on the SMD workstation.
5. Open an existing workspace by selecting
Load Workspace
in the
File
menu or create a
new workspace by selecting
New Workspace
in the
File
menu (see SymPhoTime
manual).
6. Set the correct fluorescence filter on the corresponding filter holder position of the
external PicoQuant detector unit or on the Leica TCS SP8. Select the correct SMD filter
cube in the X1 port adapter (
). If you have a dual-channel APD
detector unit, set the variable beam splitter unit to the correct position (see dual channel
detector unit manual for details).
7. Be sure that no light falls on the detectors. Either switch on the power supply or, if using
a DSN 102 power supply, start the respective detector by selecting the corresponding
channel button on the DSN 102.
The SMD upgrade is now ready to use. The communication between the SP8 confocal
system itself and the SMD upgrade electronics starts at the beginning of the data
acquisition.
Summary of Contents for TCS SP8 SMD
Page 1: ...10 Living up to Life User Manual Leica TCS SP8 SMD for FCS FLIM and FLCS ...
Page 4: ...4 Copyright ...
Page 14: ...14 Contents ...
Page 18: ...18 Intended Use ...
Page 20: ...20 Liability and Warranty ...
Page 28: ...28 General Safety Notes ...
Page 32: ...32 Additional Notes on Handling the System ...
Page 44: ...44 System Overview and Properties ...
Page 60: ...60 SMD Components Figure 31 DSN 102 Dual SPAD Power Supply ...
Page 80: ...80 Safety Features ...
Page 102: ...102 Switching On the System ...
Page 116: ...116 LAS AF ...
Page 214: ...214 Changing the Specimen ...
Page 216: ...216 Changing the Objective ...
Page 238: ...238 Switching Off the System ...
Page 242: ...242 Repairs and Service Work ...
Page 244: ...244 Maintenance ...
Page 246: ...246 Disassembly and Transport ...
Page 248: ...248 Disposal ...
Page 254: ...254 Contact ...
Page 256: ...256 Recommended Literature ...
Page 266: ...266 Appendix Figure 225 Declaration of conformity ...
Page 268: ...268 Appendix ...
Page 269: ......