F(L)CS Data Acquisition
Figure 176:
Starting the FCS test and applying the measuring point
5. To end illumination of the specimen, click the
Stop FCS Test
button in LAS AF. This will
also stop the online correlation display in SymPhoTime.
6. If you would like to reuse a tested FCS measuring point later for the 'correct' FCS
measurement, click
Add Point
) in the
Setup
tab. This measuring point
is then written to a position list, which is used in the
Measurements
step in LAS AF. Do
not make any more changes to the zoom. Changing the zoom in the
Setup FCS
step would
render the position list invalid.
The instrument parameters set for the FCS measurements (laser lines and associated
intensity, pinhole, AOBS) are kept as separate FCS settings that are independent from the
image acquisition setting defined in the
Setup Imaging
step. The FCS settings defined in the
Setup FCS
step are automatically transferred to the
Measurements
step.
18.3.6
Count Rate Monitor
The count rate of each channel is displayed in the
Setup
tab. For FCS, the count rate should
be between 10,000 and 500,000 cps (counts per second).
If the two detection channels are reversed, change the allocation in SymPhoTime. The
allocation of the channels is preset and should not be changed.
• FCS Channel A: Check box
1
• FCS Channel B: Check box
2
Saving the data in Measurements
This curve is not saved. If you want to save your data, you have to work in
the
Measurements
step.
Summary of Contents for TCS SP8 SMD
Page 1: ...10 Living up to Life User Manual Leica TCS SP8 SMD for FCS FLIM and FLCS ...
Page 4: ...4 Copyright ...
Page 14: ...14 Contents ...
Page 18: ...18 Intended Use ...
Page 20: ...20 Liability and Warranty ...
Page 28: ...28 General Safety Notes ...
Page 32: ...32 Additional Notes on Handling the System ...
Page 44: ...44 System Overview and Properties ...
Page 60: ...60 SMD Components Figure 31 DSN 102 Dual SPAD Power Supply ...
Page 80: ...80 Safety Features ...
Page 102: ...102 Switching On the System ...
Page 116: ...116 LAS AF ...
Page 214: ...214 Changing the Specimen ...
Page 216: ...216 Changing the Objective ...
Page 238: ...238 Switching Off the System ...
Page 242: ...242 Repairs and Service Work ...
Page 244: ...244 Maintenance ...
Page 246: ...246 Disassembly and Transport ...
Page 248: ...248 Disposal ...
Page 254: ...254 Contact ...
Page 256: ...256 Recommended Literature ...
Page 266: ...266 Appendix Figure 225 Declaration of conformity ...
Page 268: ...268 Appendix ...
Page 269: ......