![Leica TCS SP8 SMD User Manual Download Page 48](http://html1.mh-extra.com/html/leica/tcs-sp8-smd/tcs-sp8-smd_user-manual_1873063048.webp)
48
Ambient Conditions
8.4.2
System with Compact Supply Unit (Only with FLIM)
Supply voltage
100 V~ to 240 V~ ± 10%, grounded
Frequency
50/60 Hz
Power connection
Two separate electric circuits with
Power supply voltage: 100 V - 120 V, fuse protection: 20 A
or:
Power supply voltage: 200 V - 240 V, fuse protection: 12 - 16 A
(One for the compact supply unit's power supply and one for
supplying peripheral devices)
To prevent damage to the instrument, the two types of
voltage (200 to 240 V~ and 100 to 120 V~) must not be
combined. Depending on the system configuration, further
electrical circuits may be required.
Power supply for USA
Terminals of type NEMA 5-15
for the SMD rack
a separate electric circuit with
Power supply voltage: 100 V~ to 240 V~, fuse protection: 16 -
20 A, dependent on local regulations
for the infrared laser and its
cooling system (MP
configuration)
a separate electric circuit with
Power supply voltage: 100 V~ to 240 V~, fuse protection: 12 -
16 A, dependent on local regulations
When using HyD RLDs
An additional power supply
8.4.3
External Lasers
Observe the user manuals for external lasers
Please refer to the information from the documents provided by the laser
manufacturer for the external lasers. Pay particular attention to the laser
manufacturer's notes!
Summary of Contents for TCS SP8 SMD
Page 1: ...10 Living up to Life User Manual Leica TCS SP8 SMD for FCS FLIM and FLCS ...
Page 4: ...4 Copyright ...
Page 14: ...14 Contents ...
Page 18: ...18 Intended Use ...
Page 20: ...20 Liability and Warranty ...
Page 28: ...28 General Safety Notes ...
Page 32: ...32 Additional Notes on Handling the System ...
Page 44: ...44 System Overview and Properties ...
Page 60: ...60 SMD Components Figure 31 DSN 102 Dual SPAD Power Supply ...
Page 80: ...80 Safety Features ...
Page 102: ...102 Switching On the System ...
Page 116: ...116 LAS AF ...
Page 214: ...214 Changing the Specimen ...
Page 216: ...216 Changing the Objective ...
Page 238: ...238 Switching Off the System ...
Page 242: ...242 Repairs and Service Work ...
Page 244: ...244 Maintenance ...
Page 246: ...246 Disassembly and Transport ...
Page 248: ...248 Disposal ...
Page 254: ...254 Contact ...
Page 256: ...256 Recommended Literature ...
Page 266: ...266 Appendix Figure 225 Declaration of conformity ...
Page 268: ...268 Appendix ...