FLIM Data Acquisition
10. Define the duration of the individual FLIM images as described in
. If all
images are to have a comparable brightness, use the
Acquire until max ___ photons/
pixel
option. It enables corrections of intensity changes which are caused by the
excitation spectrum and by bleaching artefacts. If intensity information is required, use
Duration of each image
or
Repetitions
.
11. Start the FLIM stack acquisition with the image acquisition button
Run FLIM
.
In LAS AF, you obtain a
Λ
stack of time-averaged images and, in SymPhoTime, a series of
FLIM images.
16.3.10
Control of FLIM Measurements
The measurement series is started by clicking on the
Start FLIM
image acquisition button. It
stops automatically after all measurements are taken. A user-defined stop is possible by
clicking the
Stop FLIM
image acquisition button.
In the count rate monitor (
), the current count rate (in cps - counts per second) is
displayed during the individual measurements. In the line
Max. counts per pixel
, the total
number of photons accumulated so far in the brightest pixel of the image is shown. This
gives an initial indication as to whether the photon statistics in the image are sufficient.
After start of the FLIM series, fast FLIM images are displayed in SymPhoTime online and
saved together with the raw data.
Both the count rate and fast FLIM images give a first impression about the quality of the
measurements.
Figure 147:
Count rate monitor in the Measurements step of the FLIM Wizard
Summary of Contents for TCS SP8 SMD
Page 1: ...10 Living up to Life User Manual Leica TCS SP8 SMD for FCS FLIM and FLCS ...
Page 4: ...4 Copyright ...
Page 14: ...14 Contents ...
Page 18: ...18 Intended Use ...
Page 20: ...20 Liability and Warranty ...
Page 28: ...28 General Safety Notes ...
Page 32: ...32 Additional Notes on Handling the System ...
Page 44: ...44 System Overview and Properties ...
Page 60: ...60 SMD Components Figure 31 DSN 102 Dual SPAD Power Supply ...
Page 80: ...80 Safety Features ...
Page 102: ...102 Switching On the System ...
Page 116: ...116 LAS AF ...
Page 214: ...214 Changing the Specimen ...
Page 216: ...216 Changing the Objective ...
Page 238: ...238 Switching Off the System ...
Page 242: ...242 Repairs and Service Work ...
Page 244: ...244 Maintenance ...
Page 246: ...246 Disassembly and Transport ...
Page 248: ...248 Disposal ...
Page 254: ...254 Contact ...
Page 256: ...256 Recommended Literature ...
Page 266: ...266 Appendix Figure 225 Declaration of conformity ...
Page 268: ...268 Appendix ...
Page 269: ......