146
FLIM Data Acquisition
The following window is displayed:
Figure 130:
Oscilloscope window in SymPhoTime
4. In the
Acquisition
field, you can change the resolution under
resol.
Please confirm the
settings by selecting
apply
. To cover the whole range between two consecutive laser
pulses at the maximum resolution possible, the following settings are recommended:
Table 20: Recommended resolution for data acquisition with different laser frequencies
The default is a setting of 16 ps suitable for a laser repetition rate down to 20 MHz.
16.2.5
2-Laser PIE (405, 470, 640 nm)
PIE (pulsed interleaved excitation) means that two lasers pulse alternatingly. This method is
used to identify crosstalk and cross excitation and remove them from the signal. In the SMD
system, this option is available for certain combinations of pulsed diode lasers (405+470,
405+640, 470+640). Exactly two diode lasers have to be selected in order to activate 2-laser
PIE. If this is not the case, the corresponding radio button is grayed out.
Laser frequency
Resolution of data acquisition
80 MHz (MP laser, white light laser)
4 ps
40 MHz (pulsed diode laser, white light laser)
8 ps
20 MHz (pulsed diode laser, white light laser)
16 ps
10 MHz (pulsed diode laser, white light laser)
32 ps
5 MHz (pulsed diode laser)
64 ps
Summary of Contents for TCS SP8 SMD
Page 1: ...10 Living up to Life User Manual Leica TCS SP8 SMD for FCS FLIM and FLCS ...
Page 4: ...4 Copyright ...
Page 14: ...14 Contents ...
Page 18: ...18 Intended Use ...
Page 20: ...20 Liability and Warranty ...
Page 28: ...28 General Safety Notes ...
Page 32: ...32 Additional Notes on Handling the System ...
Page 44: ...44 System Overview and Properties ...
Page 60: ...60 SMD Components Figure 31 DSN 102 Dual SPAD Power Supply ...
Page 80: ...80 Safety Features ...
Page 102: ...102 Switching On the System ...
Page 116: ...116 LAS AF ...
Page 214: ...214 Changing the Specimen ...
Page 216: ...216 Changing the Objective ...
Page 238: ...238 Switching Off the System ...
Page 242: ...242 Repairs and Service Work ...
Page 244: ...244 Maintenance ...
Page 246: ...246 Disassembly and Transport ...
Page 248: ...248 Disposal ...
Page 254: ...254 Contact ...
Page 256: ...256 Recommended Literature ...
Page 266: ...266 Appendix Figure 225 Declaration of conformity ...
Page 268: ...268 Appendix ...
Page 269: ......