148
FLIM Data Acquisition
Figure 134:
Selecting the white light laser pulse frequency in Beam Path Settings
If the repetition rate is changed, it can become necessary to adapt the resolution in
SymPhoTime (see
16.2.7
Setting the Pinhole
On the
Acquisition
tab in LAS AF, set the pinhole diameter (
Pinhole
) to
1 Airy
for FLIM
measurements with VIS and UV excitation and to the maximum value for MP excitation.
16.2.8
Optimizing FLIM Settings
1. First, find the position of your specimen you want to use for FLIM measurement.
2. Click on the image acquisition button
Run FLIM Test
to start the scanning operation. You
can view and optimize the fluorescence intensity in the
Count Rate Monitor
on the
Setup
tab (
3. Laser line and intensity can be modified during the test measurement until you attain the
count rate desired.
4. When you start the test measurement, an online FLIM image is displayed in
SymPhoTime.
Do not select other frequencies on the laser
If you select
manual
in the pull-down menu, you can select additional
frequencies on the laser yourself. This is, however, not recommended
because the cable lengths are not calibrated for these frequencies.
Scan speeds 100 to 200 Hz recommended
Do not use scan speeds that are too high because these can lead to
longer analysis times in SymPhoTime. Scan speeds at 100 to 200 Hz are
recommended. Avoid the use of the resonant scanner for FLIM data
acquisition as much as possible.
The online FLIM image is not saved.
This image is not saved. To save your data, you have to work in the
Measurements
step.
Summary of Contents for TCS SP8 SMD
Page 1: ...10 Living up to Life User Manual Leica TCS SP8 SMD for FCS FLIM and FLCS ...
Page 4: ...4 Copyright ...
Page 14: ...14 Contents ...
Page 18: ...18 Intended Use ...
Page 20: ...20 Liability and Warranty ...
Page 28: ...28 General Safety Notes ...
Page 32: ...32 Additional Notes on Handling the System ...
Page 44: ...44 System Overview and Properties ...
Page 60: ...60 SMD Components Figure 31 DSN 102 Dual SPAD Power Supply ...
Page 80: ...80 Safety Features ...
Page 102: ...102 Switching On the System ...
Page 116: ...116 LAS AF ...
Page 214: ...214 Changing the Specimen ...
Page 216: ...216 Changing the Objective ...
Page 238: ...238 Switching Off the System ...
Page 242: ...242 Repairs and Service Work ...
Page 244: ...244 Maintenance ...
Page 246: ...246 Disassembly and Transport ...
Page 248: ...248 Disposal ...
Page 254: ...254 Contact ...
Page 256: ...256 Recommended Literature ...
Page 266: ...266 Appendix Figure 225 Declaration of conformity ...
Page 268: ...268 Appendix ...
Page 269: ......