207
F(L)CS Data Acquisition
This makes it easier to allocate the images in LAS AF to the corresponding FCS data in
SymPhoTime.
18.4.5
FCS z-Stack
An FCS z-stack can be particularly useful for FCS measurements on membranes. To create
an FCS z-stack, follow these steps:
1. First acquire an xyz stack in the
Setup Imaging
step in the FCS Wizard (
.
2. If necessary, optimize the measurement conditions in the
Setup FCS
step (
).
3. Change to the
Measurements
step.
4. Use the image display to find the xy point in the z image stack at which you want to
generate the FCS z-stack.
5. In the z image stack, go to the top z plane that is of interest to you. Click on the desired
xy position and then on
Add
in
Beam Path Settings
). Go to the plane below;
do not change the xy position, but click
Add
. Then go to the plane below that, etc.
Continue until you reach the lowest plane that is of interest to you.
If you wish, you can also acquire the z FCS stack from the bottom upwards.
6. Define the measurement duration, the number of repetitions at each point, and the
number of cycles as described in
. The FCS instrument parameter settings
are identical at all measuring points.
7. Specify folder and file names as described in
.
8. Click the
Run FCS
image acquisition button in LAS AF.
The maximum number of FCS measurements in a measurement series is 10,000.
Summary of Contents for TCS SP8 SMD
Page 1: ...10 Living up to Life User Manual Leica TCS SP8 SMD for FCS FLIM and FLCS ...
Page 4: ...4 Copyright ...
Page 14: ...14 Contents ...
Page 18: ...18 Intended Use ...
Page 20: ...20 Liability and Warranty ...
Page 28: ...28 General Safety Notes ...
Page 32: ...32 Additional Notes on Handling the System ...
Page 44: ...44 System Overview and Properties ...
Page 60: ...60 SMD Components Figure 31 DSN 102 Dual SPAD Power Supply ...
Page 80: ...80 Safety Features ...
Page 102: ...102 Switching On the System ...
Page 116: ...116 LAS AF ...
Page 214: ...214 Changing the Specimen ...
Page 216: ...216 Changing the Objective ...
Page 238: ...238 Switching Off the System ...
Page 242: ...242 Repairs and Service Work ...
Page 244: ...244 Maintenance ...
Page 246: ...246 Disassembly and Transport ...
Page 248: ...248 Disposal ...
Page 254: ...254 Contact ...
Page 256: ...256 Recommended Literature ...
Page 266: ...266 Appendix Figure 225 Declaration of conformity ...
Page 268: ...268 Appendix ...
Page 269: ......