FLIM Data Acquisition
Figure 122:
Selecting external HyD RLD for FLIM data acquisition
5. Deactivate the internal detectors. Internal spectral detectors cannot be used
simultaneously with HyD RLD.
6. If no image is displayed, check on the stand display whether a decoupling mirror from
the stand is in the
RLD/TLD
position. If not, set the filter cube on the stand to this position.
7. Activate at least one detector, even if no image in LAS AF is required. If no detector is
activated in LAS AF at all, it will not be possible to start a scanning operation.
Overload protection for HyD RLD
HyD RLDs are temporarily deactivated when emission intensity gets too
high. They switch back on automatically the next time a scan starts.
Beforehand, reduce the incident light on the detector.
Dependency of the spectral detection range of the HyD RLD
The spectral detection range of the HyD RLD is determined by the NDD
filter cube used, not by the slider settings for the photomultipliers.
Summary of Contents for TCS SP8 SMD
Page 1: ...10 Living up to Life User Manual Leica TCS SP8 SMD for FCS FLIM and FLCS ...
Page 4: ...4 Copyright ...
Page 14: ...14 Contents ...
Page 18: ...18 Intended Use ...
Page 20: ...20 Liability and Warranty ...
Page 28: ...28 General Safety Notes ...
Page 32: ...32 Additional Notes on Handling the System ...
Page 44: ...44 System Overview and Properties ...
Page 60: ...60 SMD Components Figure 31 DSN 102 Dual SPAD Power Supply ...
Page 80: ...80 Safety Features ...
Page 102: ...102 Switching On the System ...
Page 116: ...116 LAS AF ...
Page 214: ...214 Changing the Specimen ...
Page 216: ...216 Changing the Objective ...
Page 238: ...238 Switching Off the System ...
Page 242: ...242 Repairs and Service Work ...
Page 244: ...244 Maintenance ...
Page 246: ...246 Disassembly and Transport ...
Page 248: ...248 Disposal ...
Page 254: ...254 Contact ...
Page 256: ...256 Recommended Literature ...
Page 266: ...266 Appendix Figure 225 Declaration of conformity ...
Page 268: ...268 Appendix ...
Page 269: ......