194
F(L)CS Data Acquisition
For more detailed information about using the pulsed UV laser, refer to
18.2.2.5
Using White Light Lasers
1. Select the desired laser lines in the
Whitelight
dialog in
Beam Path Settings
.
2. Make sure that the AOBS is at fluorescence.
3. Set the desired intensity using the AOTF slider.
4. If the system has a multifunction port (
MFP
), it should be set to
Substrate
.
5. Open the
Acquisition
tab and adjust the rotation to 0° using the slider (
).
Figure 171:
Adjusting the diameter of the pinhole and rotation
18.2.3
Adjusting the Pinhole for Image Acquisition
For VIS and UV excitation, set the
Pinhole
diameter to
1 Airy
; for MP excitation set it to the
maximum size (
).
Positioning the multifunction port
Within the FCS Wizard, positioning of the multifunction port is automated;
outside the wizard, it must be set by the user. The automatic
configuration can be changed subsequently by the user.
Summary of Contents for TCS SP8 SMD
Page 1: ...10 Living up to Life User Manual Leica TCS SP8 SMD for FCS FLIM and FLCS ...
Page 4: ...4 Copyright ...
Page 14: ...14 Contents ...
Page 18: ...18 Intended Use ...
Page 20: ...20 Liability and Warranty ...
Page 28: ...28 General Safety Notes ...
Page 32: ...32 Additional Notes on Handling the System ...
Page 44: ...44 System Overview and Properties ...
Page 60: ...60 SMD Components Figure 31 DSN 102 Dual SPAD Power Supply ...
Page 80: ...80 Safety Features ...
Page 102: ...102 Switching On the System ...
Page 116: ...116 LAS AF ...
Page 214: ...214 Changing the Specimen ...
Page 216: ...216 Changing the Objective ...
Page 238: ...238 Switching Off the System ...
Page 242: ...242 Repairs and Service Work ...
Page 244: ...244 Maintenance ...
Page 246: ...246 Disassembly and Transport ...
Page 248: ...248 Disposal ...
Page 254: ...254 Contact ...
Page 256: ...256 Recommended Literature ...
Page 266: ...266 Appendix Figure 225 Declaration of conformity ...
Page 268: ...268 Appendix ...
Page 269: ......