![Leica TCS SP8 SMD User Manual Download Page 185](http://html1.mh-extra.com/html/leica/tcs-sp8-smd/tcs-sp8-smd_user-manual_1873063185.webp)
185
F(L)CS Data Acquisition
6. Go to a higher zoom (e.g. 4 to 8) and activate a continuous xz scanning operation with the
Live
image acquisition button.
7. Rotate the correction ring into a position that provides the highest contrast in the
reflection between the cover slip and the specimen (greatest possible intensity and
symmetry, narrowest possible intensity maximum without secondary maximum values;
and
Figure 162:
Reflection between cover slip and measurement medium with poorly set position of the
correction ring: The image is out of focus, several secondary peaks are visible
Figure 163:
Reflection between cover slip and measurement medium with correctly set position of the
correction ring: The image is sharp, and secondary peaks are scarcely visible
Reversed image with inverted microscope
Please note that with an inverted microscope the image is inverted. In
this case, the bottom reflection (between glass and specimen) is the
important signal that has to be optimized.
Summary of Contents for TCS SP8 SMD
Page 1: ...10 Living up to Life User Manual Leica TCS SP8 SMD for FCS FLIM and FLCS ...
Page 4: ...4 Copyright ...
Page 14: ...14 Contents ...
Page 18: ...18 Intended Use ...
Page 20: ...20 Liability and Warranty ...
Page 28: ...28 General Safety Notes ...
Page 32: ...32 Additional Notes on Handling the System ...
Page 44: ...44 System Overview and Properties ...
Page 60: ...60 SMD Components Figure 31 DSN 102 Dual SPAD Power Supply ...
Page 80: ...80 Safety Features ...
Page 102: ...102 Switching On the System ...
Page 116: ...116 LAS AF ...
Page 214: ...214 Changing the Specimen ...
Page 216: ...216 Changing the Objective ...
Page 238: ...238 Switching Off the System ...
Page 242: ...242 Repairs and Service Work ...
Page 244: ...244 Maintenance ...
Page 246: ...246 Disassembly and Transport ...
Page 248: ...248 Disposal ...
Page 254: ...254 Contact ...
Page 256: ...256 Recommended Literature ...
Page 266: ...266 Appendix Figure 225 Declaration of conformity ...
Page 268: ...268 Appendix ...
Page 269: ......