![Leica TCS SP8 SMD User Manual Download Page 173](http://html1.mh-extra.com/html/leica/tcs-sp8-smd/tcs-sp8-smd_user-manual_1873063173.webp)
173
Summarized manual for FLIM experiments
17.12.1.3
With Fluorescence Mode
This is the preferred mode for MP and pulsed VIS or UV excitation, because the conditions
of the IRF measurement match the measurement conditions of the real experiment.
1. Use a dye with a very short lifetime (a few picoseconds) in a measurement chamber. A
dye can be used for IRF measurements if the fluorescence decay time is much shorter
than the IRF width. An example for such a dye is picric acid.
2. For external FLIM detection select an appropriate SMD filter cube which will be used for
your final measurement. For SP FLIM detection, make sure the fluorifier disc is in the
correct position (
3. An additional attenuation filter in the filter holder in front of the external detector may not
be required.
Dyes with a very short lifetime can be made with the following recipe: A saturated aqueous
dye solution is created from a spectrally suitable dye with good water solubility. Potassium
iodide is added to this solution until the dye solution is saturated with it. The potassium
iodide usually quenches the fluorescence so strongly that the fluorescence lifetime of the
solution sinks to below 100 ps FWHM and thus can also be used for IRF measurements. The
advantage of this method is that the IRF can be measured at the same emission wavelengths
as it is in the final measurement. This method can also be used for multi-photon excitation
(refer also to Szabelski M. et al., Collisional quenching of Erythrosine B as a potential
reference dye for impulse response function evaluation, Applied Spectroscopy, 63, 2009).
17.12.1.4
With SHG (Second Harmonic Generation – Possible for MP Lasers Only)
1. Use a SHG-capable specimen without lifetime or with very short lifetime (a few
picoseconds) in a measurement chamber. A typical example of a SHG specimen is
muscle tissue, collagen or urea powder mounted on a cover slip.
2. Emission happens at roughly half the excitation wavelength (for instance excitation at
900 nm, emission at 450 nm). For external FLIM detection, select an appropriate SMD
filter cube that allows SHG emission to pass through.
3. The fluorifier disc has to be in the in position
SP700
or
SP680
.
4. For SP FLIM detectors, move the spectral slider such that the emitted light hits the SP
FLIM detector.
5. For external detection, remove all additional filters from the filter holder in front of the
detector.
WARNING
Risk of severe injuries from picric acid
Picric acid is poisonous and causes a danger of explosion. Observe your
country-specific regulations, which are indicated on the safety data
sheet.
Summary of Contents for TCS SP8 SMD
Page 1: ...10 Living up to Life User Manual Leica TCS SP8 SMD for FCS FLIM and FLCS ...
Page 4: ...4 Copyright ...
Page 14: ...14 Contents ...
Page 18: ...18 Intended Use ...
Page 20: ...20 Liability and Warranty ...
Page 28: ...28 General Safety Notes ...
Page 32: ...32 Additional Notes on Handling the System ...
Page 44: ...44 System Overview and Properties ...
Page 60: ...60 SMD Components Figure 31 DSN 102 Dual SPAD Power Supply ...
Page 80: ...80 Safety Features ...
Page 102: ...102 Switching On the System ...
Page 116: ...116 LAS AF ...
Page 214: ...214 Changing the Specimen ...
Page 216: ...216 Changing the Objective ...
Page 238: ...238 Switching Off the System ...
Page 242: ...242 Repairs and Service Work ...
Page 244: ...244 Maintenance ...
Page 246: ...246 Disassembly and Transport ...
Page 248: ...248 Disposal ...
Page 254: ...254 Contact ...
Page 256: ...256 Recommended Literature ...
Page 266: ...266 Appendix Figure 225 Declaration of conformity ...
Page 268: ...268 Appendix ...
Page 269: ......