Switching On the System
Figure 79:
Key switch on external UV laser 405
13. After the workstation has started, log on to the operating system.
Using a personalized user ID
Use your personal user ID if one has been set up. This ensures that user-
specific settings are saved and maintained for this user only. If the
system administrator has not yet assigned a personal user ID, log on as
"TCS_User". The default password is "True!Confocal55".
Change password
After logging on with your own user ID, you may change your password.
To do so, press the
Ctrl
,
Alt
and
Del
keys at the same time.
Then, click
Change password
. The
Change password
dialog box opens.
Now you can start the LAS AF software.
13.3
HyD Reflected Light Detectors (HyD RLDs)
• Never change or remove the filter cube or short pass filter during operation.
• The filter cube cover (see
1) ensures the housing's EMC stability and
prevents light from entering or exiting.
• Replacement of the short pass filter is only allowed starting from LAS AF Version 2.7 or
later.
If you are using HyD RLDs, you have to switch them on before LAS AF starts. Otherwise, the
LAS AF software cannot initialize the detectors. The HyD RLDs consist of a supply unit (see
) and a detector module (see
).
The power switch for switching on and off the power supply and the cooling for the detector
module is located on the front side of the supply unit (see
For image acquisition, activate the detectors in LAS AF. The yellow status LED (see
2) flashes if photons are being detected.
Summary of Contents for TCS SP8 SMD
Page 1: ...10 Living up to Life User Manual Leica TCS SP8 SMD for FCS FLIM and FLCS ...
Page 4: ...4 Copyright ...
Page 14: ...14 Contents ...
Page 18: ...18 Intended Use ...
Page 20: ...20 Liability and Warranty ...
Page 28: ...28 General Safety Notes ...
Page 32: ...32 Additional Notes on Handling the System ...
Page 44: ...44 System Overview and Properties ...
Page 60: ...60 SMD Components Figure 31 DSN 102 Dual SPAD Power Supply ...
Page 80: ...80 Safety Features ...
Page 102: ...102 Switching On the System ...
Page 116: ...116 LAS AF ...
Page 214: ...214 Changing the Specimen ...
Page 216: ...216 Changing the Objective ...
Page 238: ...238 Switching Off the System ...
Page 242: ...242 Repairs and Service Work ...
Page 244: ...244 Maintenance ...
Page 246: ...246 Disassembly and Transport ...
Page 248: ...248 Disposal ...
Page 254: ...254 Contact ...
Page 256: ...256 Recommended Literature ...
Page 266: ...266 Appendix Figure 225 Declaration of conformity ...
Page 268: ...268 Appendix ...
Page 269: ......