157
FLIM Data Acquisition
7. Enter the desired value in the time series dialog.
8. Start the FLIM experiment recording with the
Run FLIM
button.
In LAS AF, you obtain a time series of averaged images of stacks and, in SymPhoTime, a
series of FLIM images.
16.3.7
Defining an xy
λ
or xz
λ
FLIM Stack
These modes are only available for systems that are equipped with internal SP FLIM
detectors. During the generation of stacks, the spectral sliders of the SP FLIM detectors
are shifted incrementally. To acquire an FLIM stack, proceed as follows:
1. Select the specimen position.
2. In the
Setup FLIM
step on the
Setup
tab, select the
Internal (SP FLIM)
option
3. In the
Measurements
step on the
Acquisition
tab, select either the
Acquisition Mode
xy
λ
or
xz
λ
as the scan mode. Optimize the instrument parameter setting for FLIM
measurement.
4. Define a file name in the
Setup
tab (
).
5. Open the
Acquisition
tab. If the correct SP FLIM detector is not yet selected, select it in
the
PMT-Selection
line. In
Beam Path Settings
, the detector selection cannot be
changed.
6. Define the stack in the
λ
stack dialog as follows (
•
Detection Begin
and
Detection End
: Wavelength of the first and last FLIM image
•
Detection Band Width
: Width of the spectral range of an individual FLIM image
•
No. of Detection Steps
: Number of FLIM images that are acquired from different
spectral ranges
•
-Detection Stepsize
: Step size of the spectral slider from one image to the next
λ
λ
Summary of Contents for TCS SP8 SMD
Page 1: ...10 Living up to Life User Manual Leica TCS SP8 SMD for FCS FLIM and FLCS ...
Page 4: ...4 Copyright ...
Page 14: ...14 Contents ...
Page 18: ...18 Intended Use ...
Page 20: ...20 Liability and Warranty ...
Page 28: ...28 General Safety Notes ...
Page 32: ...32 Additional Notes on Handling the System ...
Page 44: ...44 System Overview and Properties ...
Page 60: ...60 SMD Components Figure 31 DSN 102 Dual SPAD Power Supply ...
Page 80: ...80 Safety Features ...
Page 102: ...102 Switching On the System ...
Page 116: ...116 LAS AF ...
Page 214: ...214 Changing the Specimen ...
Page 216: ...216 Changing the Objective ...
Page 238: ...238 Switching Off the System ...
Page 242: ...242 Repairs and Service Work ...
Page 244: ...244 Maintenance ...
Page 246: ...246 Disassembly and Transport ...
Page 248: ...248 Disposal ...
Page 254: ...254 Contact ...
Page 256: ...256 Recommended Literature ...
Page 266: ...266 Appendix Figure 225 Declaration of conformity ...
Page 268: ...268 Appendix ...
Page 269: ......