158
FLIM Data Acquisition
Figure 142:
Defining xy FLIM stacks
7. Define the duration of the individual FLIM images as described in
. If all
images are to have a comparable brightness, use the
Acquire until max ___ photons/
pixel
option. It enables corrections of intensity changes which are caused by the
emissions spectrum and by bleaching artefacts. If spectral intensity information is
required, use
Duration of each image
or
Repetitions
.
8. Start the FLIM stack acquisition with the image acquisition button
Run FLIM
.
In LAS AF, you obtain a
λ
stack of averaged images and, in SymPhoTime, a series of FLIM
images.
16.3.8
Defining a Time Series of xy
λ
or xz
λ
FLIM Stacks
These modes are only available for systems that are also equipped with internal SP FLIM
detectors.
1. Select the specimen position.
2. In the
Setup FLIM
step on the
Setup
tab, select the
Internal (SP FLIM)
option
3. In the
Measurements
step on the
Acquisition
tab, select either the
Acquisition Mode
xy
λ
t or
xz
λ
t as the scan mode. Optimize the instrument parameter setting for FLIM
measurement.
4. Define a file name in the
Setup
tab (
).
5. Open the
Acquisition
tab. If the correct SP FLIM detector is not yet selected, select it in
the
PMT-Selection
line. In
Beam Path Settings
, the detector selection cannot be
changed.
λ
Summary of Contents for TCS SP8 SMD
Page 1: ...10 Living up to Life User Manual Leica TCS SP8 SMD for FCS FLIM and FLCS ...
Page 4: ...4 Copyright ...
Page 14: ...14 Contents ...
Page 18: ...18 Intended Use ...
Page 20: ...20 Liability and Warranty ...
Page 28: ...28 General Safety Notes ...
Page 32: ...32 Additional Notes on Handling the System ...
Page 44: ...44 System Overview and Properties ...
Page 60: ...60 SMD Components Figure 31 DSN 102 Dual SPAD Power Supply ...
Page 80: ...80 Safety Features ...
Page 102: ...102 Switching On the System ...
Page 116: ...116 LAS AF ...
Page 214: ...214 Changing the Specimen ...
Page 216: ...216 Changing the Objective ...
Page 238: ...238 Switching Off the System ...
Page 242: ...242 Repairs and Service Work ...
Page 244: ...244 Maintenance ...
Page 246: ...246 Disassembly and Transport ...
Page 248: ...248 Disposal ...
Page 254: ...254 Contact ...
Page 256: ...256 Recommended Literature ...
Page 266: ...266 Appendix Figure 225 Declaration of conformity ...
Page 268: ...268 Appendix ...
Page 269: ......